LiebherrSRFvh 5511 Perfection
- Click here for all specifications of the Liebherr SRFvh 5511
- Click here for the user manual of the Liebherr SRFvh 5511
The 2.2" colour display (12 languages including Dutch) can also be operated with gloves. In the event of a temperature alarm, SmartLock automatically locks the cooler door so that products with an interrupted cooling chain do not accidentally come into circulation. Authorised persons can unlock the lock again. To guarantee full traceability, the integrated data logger records temperature profile, alarm history and service information for 5 years. The stored data can be read via a USB interface. Meets the requirements for storing medicines (DIN 13277)
This medical refrigerator is suitable for professional storage of products at a
temperature of 5 °C. Typical products that can be stored are refrigerated medicines, drugs or other pharmaceutical products that are stored in a temperature range of 2 °C to 8 °C
The device complies with the requirements of DIN 13277 for refrigerated and deep-freeze storage equipment for laboratory and medical applications.
When storing valuable or temperature-sensitive substances or products, the use of an independent, permanently monitored alarm system is required. This alarm system must be designed in such a way that any alarm condition is immediately registered by a responsible person,who can then take appropriate measures
The temperature inside the appliance isset to the required +5 °C and cannot be changed.
Cooling temperature setting range | +3°C to +16°C |
Maximum fluctuation cooling section | 2 °C |
Gradient cooling section | 3.3 °C |
Heating time from +5 °C to +10 °C (empty, +25 °C TU) | 60 |
Heating time from +25 °C to +5 °C (empty, +25 °C TU) | 28 |
Featured:
- Maximum positive temperature deviation from setpoint 1.2
- Maximum negative temperature deviation from setpoint - 1.7
- Recovery time after 1 min door opening (empty, +25 °C TU, according to EN 60068-3) 13
- Never defrost again thanks to the automatic defrost system
- Faster cooling of medicines through thedynamic cooling system
- The self-closing door will not remain open unnoticed.
- With left or right turning door. (interchangeable)
- Audible warning in case of malfunction through theoptical and acoustic temperature alarm
- Maximum fluctuation cooling section 2.2 °C
- Registration of alarm notifications
- Electronic lock with remote control
- Potential-free contact
- Touch & Swipe Control Panel
- Temperature easy to read via thedigital temperature display